DRY SUCTION CONTROL: The suction regulator is pre-set to –20 mm H2O. Can be adjusted to any suction setting between –10 and –40 mm. H2O. Turn the dial on the side of the drain to adjust if needed per MD Order. BELLOWS: Set suction to at least –80 mmhg or higher to expand bellows beyond the triangle indicator.
Using the Atrium Oasis Closed Chest Drain Applying Vacuum Pressure •Applying your suction: •Connect suction tubing firmly to blue suction port. •Turn up suction to full vacuum. •The chest drain requires -80mmHg vacuum pressure to function. •Turning up the suction at the source will “turn on” the drain. •Actual suction pressure is
Below, you'll find personal guides to wet and dry suction chest drainage, setup and operation videos, printable wall charts, posters, and other helpful resources. Oasis Dry Suction Water Seal Chest Drain
Manufacturer recommends a suction level of -20cmH20. The suction level is pre-set at -20cm H2O on the Atrium Drainage System. DO NOT change the setting unless prescribed by MO.
Changing suction pressure is accomplished by adju-sting the rotary dry suction control dial located on the side of the drain. Dial down to lower the suction pres-sure and dial up to increase the suction pressure.
3 days ago · By including pre-packaged sterile water to fill the water seal along with unsurpassed knock-over protection, the Oasis™ dry suction chest drain offers the ultimate in fast set-up and patient...
Atrium Oasis ™ dry suction water seal chest drains, featuring a familiar water seal operating system with the enhanced performance and convenience of dry suction control.
Suction ON: One-way valve protects the patient from atmospheric air in any orientation when suction is ON. Suction OFF: Risk of atmospheric air entering the patient around angled fluid when suction is OFF.